| Almighty anesthesia machine type:SX15 - RY - ¢ò type |
|
Technical parameters:
1. The work mode: full closed, half closed (0.5-7.5) kPa and half open
2 gas supply: can use a central piping gas supply, gas supply or cylinder pressure range: (0.27-0.55) MPa.
3. Flowmeter O2:0.1-10L/min N2O: 0.05-10L/min
4. Flow control valves: oxygen and laughing gas linkage institutions to ensure oxygen concentration no less than 25%, O2 flow control valves not open N2O cannot enter anesthesia respiratory system.
5. Enough oxygen to alarm: oxygen pressure below 0.2 MPa + 10% from time to tome voice alarm continuously and laughing gas automatically truncated time not less than 7 seconds.When oxygen pressure continued to decline, N2O flow in O2 flow stops stopped before entering system.
6. Rapid oxygen flow: 35-75L/min 7. High-precision evaporator: configuration KT - ¢ò type enflurane evaporator, optional KT - ¢ò type isoflurane evaporator, also can be equipped with imported evaporator.Concentration range: 0-5vol % in flow range: 0.25-15L/min, temperature range: 15-35 ¡æ, flow, temperature and pressure are automatically compensated.
HVJ - 880M anesthesia breathing machines
Enterprise spurt (must) words tube first 3540138 number. 2001
HVJ - 880M anesthesia breathing machines adopt pneumatic control, time switch, pressure protection function with tide volume and airway pressure will be shown, airway pressure upper and lower limit alarm.Adopt imported components and sensor, ensure parameter control precision, low gas consumption, economic and practical.Suitable for adults and children, and supporting the use of anesthesia machine.
Technical parameters:
The tidal volume: (50 ~ 1500) ml
Frequency (6 ~ 60) beats per minute
Suction than: 1:1 ~ comforteth cry
Inspiratory flow: (6 ~ 66) L/MIN
Airway pressure: (- 1 ~ 7) kPa
Pressure limit (and tent-pins) (1 ~ 6) kPa
Pressure lower limit: (0.4 ~ 4) kPa below set limit for 4 ~ 15S after alarm
The bellows: adult: (300 ~ 1500) ml pediatric: (50 ~ 300) ml |
|
|